﻿This document provides instructions for updating the RIGOL DSG3000B series RF signal generators. This firmware is intended exclusively for the DSG3000B series and is not compatible with any DSG3000 models without the ‘B’ designation. For other DSG3000 models, please visit the official RIGOL website.

[Update Requirements] 
    - Software version number: 00.01.00.00.00 and later
    - BOOT version number: 00.01.00.00.00 and later
    - Digital version number: 00.00 and later
    - RF version number: 00.00 and later
    - RF FPGA version number: 00.00 and later  
    - RF CPLD version number: 00.00 and later
    - Digital FPGA version number: 00.00.00 and later
    - Digital CPLD version number: 00.00 and later

[Notices]

    - Before updating, please review the Release Notes for detailed information about the firmware update.

    - Ensure your USB flash drive is functioning properly. The DSG3000B series require a FAT32-formatted USB flash drive, preferably 16GB or smaller.

    - Do not power off the instrument or remove the USB drive during the update process, as this may cause the device to malfunction.

[Update Procedure]

1. Press the [System] button on the front panel to open the System menu. Navigate to Information > System Info to view system details. Confirm the instrument meets the update requirements. If it does, proceed. If not, do not continue.
  
2. Extract the firmware package and copy the .sys file to the root directory of the USB drive. Do not rename the file.
  
3. With the instrument powered on, insert the USB drive into the USB port at the back of the instrument.
  
4. Press the [Storage] button on the panel to open the Storage menu. Use the knob and the [Enter] button to browse and select the firmware file.
  
5. While the firmware file is highlighted, select Update in the Storage menu.

6. The update will begin, and progress will be displayed on the screen.
  
7. Once complete, the instrument will restart automatically.
  
8. Verify that the firmware version has been successfully updated.
  
[troubleshooting]

    - If the instrument fails to read the update file:
      try using a different USB drive.

    - If the update fails due to interruption:
        a. Insert the USB drive and power on the instrument.
        b. While booting, repeatedly press the green [Preset] button (top-left of front panel) until a notification appears at the bottom of the screen showing the update progress. 
        c. Wait for the update to complete.
        d. Remove the USB drive and restart the instrument.

    - If the problem persists:
      please contact your regional support team for assistance.

· Africa, Middle East & Southern Asia: info.in@rigol.com
· Canada & the US: help@rigol.com
· China: info.cn@rigol.com
· Europe: info-europe@rigol.com
· Japan: info.jp@rigol.com
· Korea: info.kr@rigol.com
· South America: info.br@rigol.com
· All Other Regions: info.int@rigol.com